KNOWLEDGE SUMMARY
Keywords: BLOOD PRESSURE; DIAGNOSTIC TEST EVALUATION; DOPPLER; FELINE; HYPERTENSION; OSCILLOMETRIC
Is non-invasive blood pressure comparable to the invasive method in cats?
Fiona Douglas, BVMS CertAVP(VA) MRCVS1*
1 Newcastle upon Tyne, United Kingdom
* Corresponding author email: fiona.douglas@outlook.com
Vol 11, Issue 2 (2026)
Submitted 16 August 2024; Published: 24 Jun 2026
DOI: https://doi.org/10.18849/ve.v11i2.735
PICO question
In cats, is blood pressure measurement using non-invasive methods (Doppler or oscillometric) comparable to invasive arterial blood pressure measurement (direct arterial telemetry)?
Clinical bottom line
Category of research
Diagnosis.
Number and type of study designs reviewed
Thirteen method comparison studies and one prospective experimental study (fourteen total).
Strength of evidence
Moderate.
Outcomes reported
For oscillometry: two studies found it to be comparable; five studies found it to only be comparable in certain circumstances; and four studies found it not to be comparable (11 studies total measuring oscillometric). For Doppler: no studies found it to be completely comparable; two studies found it to be comparable under certain circumstances; and five found it not to be comparable (seven studies total measuring Doppler).
Conclusion
The evidence is mixed and suggests that Doppler in cats is not comparable to the reference standard (invasive measurement) or may only be comparable under certain conditions. Oscillometry may be comparable to the reference standard but is also often found to only be comparable under specific conditions. There is more evidence for the comparability of oscillometry than for Doppler, particularly for measurement of mean arterial pressure.
How to apply this evidence in practice
The application of evidence into practice should take into account multiple factors, not limited to: individual clinical expertise, patient’s circumstances and owners’ values, country, location or clinic where you work, the individual case in front of you, the availability of therapies and resources.
Knowledge Summaries are a resource to help reinforce or inform decision making. They do not override the responsibility or judgement of the practitioner to do what is best for the animal in their care.
Clinical scenario
Assessing blood pressure is important in anaesthetised or critically ill cats to monitor for hypotension, and for older cats at risk of hypertension due to diseases such as chronic kidney disease or hyperthyroidism. Senior veterinary staff or practice managers may wish to know the accuracy of Doppler or oscillometric devices when making purchasing decisions.
In some clinical scenarios, the accuracy of a device may be less important than its internal consistency in order to monitor trends, such as response to medication. However, in some clinical scenarios a threshold of blood pressure is used to prompt a specific action even if there are no previous measurements for that patient, such as the criteria for beginning anti-hypertensive treatment in cats (Taylor et al., 2017) or treating hypotension under anaesthesia (Grubb et al., 2020). Therefore, it can still be important to know how closely the non-invasive blood pressure measurements match the invasive arterial method, which is the most accurate method available.
The evidence
Thirteen method comparisons (diagnostic test evaluation studies) were analysed along with one prospective clinical study. Each compared a non-invasive method (Doppler or oscillometry) to the reference standard (direct arterial telemetry, also known as invasive). Three of these studies investigated only Doppler measurement (Cerejo et al., 2020, da Cunha et al., 2014; De Lombaert et al., 2023); seven investigated oscillometric measurement (Cerejo et al., 2017; Martel et al., 2013; Branson et al., 1997; Pedersen et al., 2002; Zwijnenberg et al., 2011; Acierno et al., 2010; Cremer et al., 2020); and four investigated both methods (Anjos et al., 2014; Binns et al., 1995; Caulkett et al., 1998; Haberman et al., 2006).
For Doppler measurement there was evidence suggesting that it is not comparable in cats, and some evidence suggesting it is only comparable under certain circumstances, particularly regarding the location and size of the cuff. For oscillometric there is some evidence that it is comparable, and other evidence that suggests it is only comparable for mean arterial pressure and not systolic pressure. Overall, the evidence is mixed and cannot substantiate that non-invasive blood pressure is accurate compared to the invasive method in cats, although there was more evidence for the comparability of oscillometric measurements than Doppler. This contrasts with the views of veterinarians; when surveyed, 69% of veterinarians believed Doppler to be more “trustworthy” than oscillometric measurements (Navarro et al., 2022). No change in clinical protocols would be suggested based on this evidence other than to take the cuff size and site of measurement into consideration when taking blood pressure measurements.
Summary of the evidence
Acierno et al. (2010)
Agreement between directly measured blood pressure and pressures obtained with three veterinary-specific oscillometric units in cats
Aim: To compare oscillometric blood pressure to directly measured blood pressure.
Population: |
Feline patients presenting for neutering to a Sterilisation Assistance Programme in the USA. |
|---|---|
Sample size: |
21 cats. |
Intervention details: |
|
Study design: |
Method comparison study. |
Outcome Studied: |
Objective measurement of oscillometric blood pressure compared to directly measured. |
Main Findings |
|
Limitations: |
|
Anjos et al. (2014)
Evaluation and comparison between methods of measurement of systolic blood pressure in healthy anesthetized cats
Aim: To compare two methods of non-invasive systolic blood pressure (Doppler and oscillometric) to invasive blood pressure.
Population: |
Healthy client-owned cats in Brazil. |
|---|---|
Sample size: |
12 cats. |
Intervention details: |
|
Study design: |
Method comparison study (experimental). |
Outcome Studied: |
|
Main Findings |
|
Limitations: |
|
Binns et al. (1995)
Doppler Ultrasonographic, Oscillometric Sphygmomanometric, and Photoplethysmographic Techniques for Noninvasive Blood Pressure Measurement in Anesthetized Cats
Aim: To compare invasive blood pressure with three non-invasive methods in anaesthetised cats.
Population: |
Adult mixed-breed cats at University of Illinois, Department of Veterinary Clinical Medicine, USA. |
|---|---|
Sample size: |
11 cats. |
Intervention details: |
|
Study design: |
Diagnostic test evaluation study. |
Outcome Studied: |
Comparison between invasive blood pressure and three non-invasive methods in anaesthetised cats. |
Main Findings |
|
Limitations: |
|
Branson et al. (1997)
Evaluation of an oscillometric blood pressure monitor on anesthetized cats and the effect of cuff placement and fur on accuracy
Aim: To determine the accuracy of a specific oscillometric machine compared to invasive blood pressure, and the effect of measurement site and fur clipping on the accuracy.
Population: |
Adult mixed-breed shorthair cats at the University of Missouri, USA. |
|---|---|
Sample size: |
6 cats. |
Intervention details: |
|
Study design: |
Method comparison study (experimental). |
Outcome Studied: |
The accuracy of a specific oscillometric machine compared to IBP, and the effect of measurement site and fur clipping on the accuracy of the measurement. |
Main Findings |
|
Limitations: |
|
Caulkett et al. (1998)
A Comparison of Indirect Blood Pressure Monitoring Techniques in the Anesthetized Cat
Aim: To measure agreement between three methods of non-invasive blood pressure measurement and the invasive method in cats.
Population: |
Adult cats at University of Saskatchewan, Canada. |
|---|---|
Sample size: |
8 cats. |
Intervention details: |
|
Study design: |
Method comparison study (experimental). |
Outcome Studied: |
Objective measurement of agreement between three methods of non-invasive blood pressure measurement and the invasive method. |
Main Findings |
|
Limitations: |
|
Cerejo et al. (2017)
Comparison of two species-specific oscillometric blood pressure monitors with direct blood pressure measurement in anesthetized cats
Aim: To compare measurement of blood pressure via two different oscillometric machines to invasive blood pressure.
Population: |
Client-owned adult cats presented for neutering at a veterinary teaching hospital in the USA. |
|---|---|
Sample size: |
8 cats. |
Intervention details: |
|
Study design: |
Method comparison study (experimental). |
Outcome Studied: |
|
Main Findings |
|
Limitations: |
|
Cerejo et al. (2020)
Effects of cuff size and position on the agreement between arterial blood pressure measured by Doppler ultrasound and through a dorsal pedal artery catheter in anesthetized cats
Aim: To determine the effect of cuff size and cuff location on agreement between Doppler measurement and invasive blood pressure.
Population: |
Client-owned cats presented for neutering at veterinary teaching hospital in Brazil. |
|---|---|
Sample size: |
8 cats. |
Intervention details: |
|
Study design: |
Method comparison study (experimental). |
Outcome Studied: |
Effect of cuff size and location on agreement between Doppler measurement and invasive blood pressure. |
Main Findings |
|
Limitations: |
|
Cremer et al. (2020)
Validation of the oscillometric blood pressure monitor Vet20 Suntech in anesthetized healthy cats
Aim: To compare oscillometric blood pressure to invasive measurements, and to attempt to validate a specific oscillometric device (Vet20 machine, Suntech Medical).
Population: |
Shelter cats presenting for neutering at Louisiana State University, USA. |
|---|---|
Sample size: |
33 cats. |
Intervention details: |
|
Study design: |
Method comparison study. |
Outcome Studied: |
Objective measurement of oscillometric blood pressure compared to invasive measurements, to attempt to validate a specific machine. |
Main Findings |
|
Limitations: |
|
da Cunha et al. (2014)
Measuring level of agreement between values obtained by directly measured blood pressure and ultrasonic Doppler flow detector in cats
Aim: To measure the agreement between Doppler blood pressure measurement and two sites of invasive blood pressure measurement.
Population: |
Adult shelter cats presented for elective sterilisation at Louisiana State University Animal Sterilization Assistance Program. |
|---|---|
Sample size: |
39 cats. |
Intervention details: |
|
Study design: |
Diagnostic test evaluation study. |
Outcome Studied: |
Objective measurement of agreement between Doppler blood pressure measurement and 2 sites of invasive blood pressure measurement. |
Main Findings |
|
Limitations: |
|
De Lombaert et al. (2023)
Effect of gabapentin on ambulatory, direct, systemic arterial blood pressure in apparently healthy cats in the at-home and in-clinic environments
Aim: To measure the effect of gabapentin on direct arterial blood pressure in cats at home and in a veterinary clinical environment.
Population: |
Purpose-bred male neutered domestic shorthair cats between 2 and 3 years of age at University of Georgia, USA. |
|---|---|
Sample size: |
5 cats (one excluded after an initial 6). |
Intervention details: |
|
Study design: |
Prospective crossover experimental study. |
Outcome Studied: |
Effect of gabapentin on direct arterial blood pressure in cats at home and in a veterinary clinical environment. |
Main Findings |
|
Limitations: |
|
Haberman et al. (2006)
Evaluation of Doppler ultrasonic and oscillometric methods of indirect blood pressure measurement in cats
Aim: To measure Doppler and oscillometric blood pressure compared to invasive blood pressure in conscious and anaesthetised cats.
Population: |
Adult mixed breed laboratory cats that were health tested, at University of Georgia (USA). |
|---|---|
Sample size: |
13 cats. |
Intervention details: |
|
Study design: |
Method comparison study. |
Outcome Studied: |
Objective measurement of Doppler and oscillometric blood pressure compared to invasive blood pressure in conscious and anaesthetised cats. |
Main Findings |
|
Limitations: |
|
Martel et al. (2013)
Comparison of high-definition oscillometry — a non-invasive technology for arterial blood pressure measurement — with a direct invasive method using radio-telemetry in awake healthy cats
Aim: To compare high definition oscillometry to invasive telemetry in conscious cats.
Population: |
Purpose-bred experimental juvenile cats in Europe. |
|---|---|
Sample size: |
6 cats. |
Intervention details: |
|
Study design: |
Method comparison study (experimental). |
Outcome Studied: |
Comparison between objective measurements of blood pressure by high definition oscillometry and invasive telemetry in conscious cats. |
Main Findings |
|
Limitations: |
|
Pedersen et al. (2002)
Evaluation of an oscillometric blood pressure monitor for use in anesthetized cats
Aim: To measure the accuracy of an oscillometric device in anaesthetised cats.
Population: |
Persian cats with subclinical polycystic kidney disease at The Royal Veterinary and Agricultural University, Frederiksberg, Denmark. |
|---|---|
Sample size: |
6 cats. |
Intervention details: |
|
Study design: |
Method comparison study. |
Outcome Studied: |
An objective measurement of the accuracy of an oscillometric device in anaesthetised cats. |
Main Findings |
|
Limitations: |
|
Zwijnenberg et al. (2011)
Evaluation of oscillometric and vascular access port arterial blood pressure measurement techniques versus implanted telemetry in anesthetized cats
Aim: To compare blood pressure measured by non-invasive oscillometry and a semi-invasive vascular access port with invasive telemetry in cats.
Population: |
Healthy domestic shorthair cats aged from 12 to 17 months in Australia. |
|---|---|
Sample size: |
6 cats. |
Intervention details: |
|
Study design: |
Method comparison study. |
Outcome Studied: |
Objective comparison of blood pressure measured by non-invasive oscillometry and a semi-invasive vascular access port with invasive telemetry in cats. |
Main Findings |
|
Limitations: |
|
Appraisal, application and reflection
This Knowledge Summary evaluates two test methods of measuring blood pressure (oscillometric and Doppler) against the method that is considered the most accurate method available (invasive arterial pressure), even if no method is truly 100% accurate. For this reason, the word comparable is mostly used in the summary. Where the term accuracy is used (for example, because the study or author being referred to also used the term) it refers to the accuracy of the device compared to the reference method (invasive method).
No systematic review articles were found in the literature search for this topic, which would be the highest level of evidence. All the articles in this Knowledge Summary except one are Level 1B (a validating cohort study) which is the next best evidence level for a study investigating diagnosis (Howick et al., 2009) The remaining study (De Lombaert et al., 2023) is an exploratory cohort study, which is Level 2 evidence equivalent to a randomised controlled trial (Howick et al., 2009).
The American College of Veterinary Internal Medicine (ACVIM) published guidelines for validation of a non-invasive blood pressure device in animals (Brown et al., 2007), which are based on the American Association of Medical Instrumentation (AAMI) criteria but less stringent and more appropriate for devices in veterinary patients. The only more recent guidance published by the AAMI in conjunction with other relevant organisations is for human devices exclusively (Stergiou et al., 2018), therefore the ACVIM standards were the most appropriate. The main standard is that the mean bias between the methods should be ≤ 10 mmHg and standard deviation (SD) ≤ 15 mmHg, which is the basis of accuracy assessment in this summary. None of the papers in this summary found the non-invasive device met all the standards for validation of the method, but some met this main standard. One paper (Cremer et al., 2020) found that the non-invasive oscillometric device met 5 of 6 standards set by ACVIM, only failing for correlation which was considered to be of least importance.
Of the seven papers assessing the Doppler method (Cerejo et al., 2020; Anjos et al., 2014; da Cunha et al., 2014, ; De Lombaert et al., 2023; Binns et al., 1995; Caulkett et al., 1998; Haberman et al., 2006), five found it to be not comparable (Anjos et al., 2014 ; da Cunha et al., 2014; De Lombaert et al., 2023; Binns et al., 1995; Haberman et al., 2006) and two found it to be comparable only under certain circumstances (Cerejo et al., 2020, and Caulkett et al., 1998). For example, Cerejo et al. (2020) found that systolic arterial pressure (SAP) by Doppler measurement met ACVIM standards (mean bias ≤ 10 mmHg and SD ≤ 15 mmHg) for agreement with invasive methods only with cuff sizes 1 and 2 on the thoracic limb and cuff size 2 placed above the tarsus. It has been suggested that in cats, measurement of SAP by Doppler may actually be a closer estimate of mean arterial pressure (MAP) because it underestimates SAP (Caulkett et al., 1998). However, this summary did not substantiate that finding as two authors found this to be untrue (De Lombaert et al. 2023; Cerejo et al., 2020).
Plethysmography was not reviewed in this summary as it is currently an uncommonly used non-invasive method in general veterinary practice (Skelding & Valverde, 2020).
Of the eleven papers assessing the oscillometric method (Cerejo et al., 2017; Anjos et al., 2014; Martel et al., 2013;Binns et al., 1995;Caulkett et al., 1998; Branson et al., 1997; Pedersen et al., 2002; Haberman et al., 2006; Zwijnenberg et al., 2011; Acierno et al., 2010; Cremer et al., 2020) two found it to be comparable (Anjos et al., 2014 and Cremer et al., 2020) and four found it to be not comparable (Binns et al., 1995; Caulkett et al., 1998; Branson et al., 1997; Haberman et al. 2006). However, five found it to be comparable only under certain circumstances (Cerejo et al., 2017; Martel et al., 2013; Pedersen et al., 2002; Zwijnenberg et al., 2011; Acierno et al., 2010). In these circumstances, Cerejo et al. (2017) found oscillometry only to be accurate for MAP and not SAP. Martel et al. (2013) found oscillometry to be not comparable during hypotension and for diastolic pressure. In contrast, Pedersen et al. (2002) found that SAP measured by oscillometry was comparable only during hypotension, while MAP was consistently comparable. Zwijnenberg et al. (2011) found that oscillometric measurements were not comparable for systolic or diastolic pressure, and only comparable for MAP.
Of the four papers that evaluated the Doppler and oscillometric method (Anjos et al., 2014; Binns et al., 1995; Caulkett et al., 1998; Haberman et al. 2006), one (Anjos et al., 2014) found that although Doppler measurements were not comparable, they were highly correlated, such that a correction factor could be applied to the results allowing Doppler to be used in a manner that would be more similar to the invasive method. These findings were not replicated by da Cunha et al. (2014) who found Doppler measurements to be poorly correlated.
There are several limitations to the studies discussed in this summary. Firstly, the cat populations and their source was varied. They consisted of laboratory cats (Martel et al., 2013; de Lombaert et al., 2023 Binns et al., 1995; Branson et al., 1997; Haberman et al., 2006; Zwijnenberg et al., 2011) which would be considered convenience sampling, shelter cats (da Cunha et al., 2014; Cremer et al., 2020), pet cats presented for neutering (Cerejo et al., 2017; Cerejo et al. 2020; Acierno et al., 2010) which is a type of consecutive sampling, tutor-owned cats (Anjos et al., 2014), ex breeding cats presented for euthanasia (Pedersen et al., 2002), and adult cats with no further specification (Caulkett et al., 1998). Most were young and considered healthy by physical exam, with or without health screening, creating a skewed sample that does not fully represent the pet cat population.
Secondly, several different arteries were used to measure both invasive and non-invasive blood pressure used across the papers. The study that investigated this specifically (Cerejo et al., 2020) found that the location of the non-invasive measurement affected the comparability of the readings significantly; any measurements taken in the hindlimb below the tarsus did not meet ACVIM standards for accuracy. The locations used for non-invasive measurement included the forelimb only in 5 papers (Anjos et al., 2014; da Cunha et al., 2014; Caulkett et al., 1998; Pedersen et al., 2002; Cremer et al., 2020), tail only in 3 papers (Martel et al., 2013; De Lombaert et al., 2023; Zwijnenberg et al., 2011), the forelimb and tail in 3 papers (Cerejo et al., 2017; Haberman et al., 2006; Acierno et al., 2010) and another combination in the remaining 3 papers (Cerejo et al. 2020; Binns et al., 1995; Branson et al., 1997).
The dorsal pedal artery, femoral artery, and abdominal aorta were all used for invasive measurement, which could achieve different results for the reference standard; this limitation was discussed by Martel et al. (2013). It has been shown in dogs (Monteiro et al., 2013) and horses (Midon et al., 2023) that results for invasive pressure from different arteries vary. However, one of the papers in this Knowledge Summary investigated both femoral and dorsal pedal arteries in different groups for invasive measurement and found no significant difference between them (da Cunha et al., 2014), so it is possible that smaller size or specific species differences mean this is not relevant to cats.
There was a great variety of equipment used in the papers analysed, including nine different multiparameter monitors for oscillometric measurement and several different cannulas and transducers. This introduces variability that is not controlled for. All seven papers evaluating the Doppler method (Anjos et al., 2014; da Cunha et al., 2014; De Lombaert et al., 2023, Binns et al., 1995; Caulkett et al., 1998; Cerejo et al., 2020; Haberman et al., 2006) used a Parks Medical Device, of which six were exactly the same model. However, only one paper of these seven was able to establish any agreement with the invasive method. All three studies that used a Dinamap device (Binns et al., 1995; Caulkett et al., 1998; Haberman et al., 2006) found oscillometry not to be comparable to invasive. In contrast, the three times a petMAP™ device was used (different specific models), it was found to agree with the invasive method for MAP every time, even if not for SAP and DAP. A Cardell device featured three times in different papers (with different specific models) and was found to be comparable for MAP in two of the three studies (Zwijnenberg et al., 2011; Pedersen et al., 2002). Other devices were only tested once across all the papers and varied in whether they could be validated as accurate compared to the invasive method. Some newer popular multiparameter devices using oscillometry such as the Mindray or Lutech Datalys did not feature in any of the studies in this summary.
Agreement between invasive measurements in different arteries can vary depending on the haemodynamic situation of the animal, such as the heart rate (Monteiro et al., 2013). This could be profoundly affected by choice of anaesthetic drugs, and the varied regimes in these papers could account for some differences in the accuracy.
Thirdly, the accuracy of non-invasive devices can vary with the blood pressure value itself (Garofalo et al., 2012), so it is important to measure the agreement with a reference standard across a range of values. Multiple papers (Anjos et al., 2014; da Cunha et al., 2014; De Lombaert et al., 2023; Acierno et al., 2010; Cremer et al., 2020) did not attempt to manipulate blood pressure to achieve a range of measures. These papers cited ethical reasons for not performing this as hypotension can cause harm and the cats were pets. Papers in this summary that did manipulate blood pressure artificially either obtained informed consent and kept the hypotensive states as short as possible or performed the experiment under terminal anaesthesia. Ethical views on this may vary between countries and institutions. There was little consistency of the method used to manipulate blood pressure, and of the ranges used to define hypo-, hyper- and normotension.
Furthermore, diagnosis of hypertension in conscious cats is an important application of BP measurement and only three papers (Martel et al., 2013; De Lombaert et al., 2023; Haberman et al., 2006) measured BP in conscious animals, only one of which was with Doppler measurements (De Lombaert et al., 2023). Since a device can only be validated for the conditions under which it was studied (Brown et al., 2007), data from eleven studies can only be applied to anaesthetised cats which is a significant limitation. Under anaesthesia it is difficult to achieve blood pressures in the hypertensive range that would mimic a clinical patient, even with pharmaceutical manipulation. Two of the papers in this summary specifically mentioned difficulty with this (Cerejo et al., 2017; Cerejo et al., 2020)) with one of them only achieving a maximum systolic pressure range of 140–150 mmHg. This does not model the hypertensive cat well, since a cat may be presented with a systolic BP of > 180 mmHg (Brown et al., 2007; Jepson, 2011). There can also be difficulty with non-invasive measurements in the hypotensive state, which is equally important to capture. Hypotension is most likely to occur under anaesthesia so assessing accuracy in anaesthetised cats is appropriate, but awake cats can still be at risk of this for other reasons, such as sepsis (Troia et al., 2019). Two studies (Cerejo et al., 2017; Caulkett et al., 1998) found the oscillometric device failed more frequently at lower invasive BP ranges. This reduces the number of valid measurements available for statistical analysis, and/or takes more time to obtain an answer. It is clinically important to recognise hypotension in the anaesthetised veterinary patient to correct it quickly, so it is equally important to be able to validate a device for use in this situation.
Replicating the methods of these studies in a population of known hypertensive cats in the conscious state would be informative but have numerous practical barriers. Inducing hypertensive diseases artificially in laboratory cats is challenging and may take significant time, but obtaining owner consent to perform anaesthesia and telemetry in pets where there are risks involved is equally difficult. Acclimatisation to the measurement to avoid white coat syndrome (situational hypertension due to handling stress) and management of comorbidities associated with hypertension for the duration of the study would also have to be considered. This could theoretically be mitigated in part by selecting cats surrendered to charities for rehoming, but this also raises ethical questions, and rescue centres may not be willing to engage with this.
The Bland-Altman method of analysis is the most appropriate for comparing two methods of quantitative measurement (Giavarina, 2015). For multiple methods in the same individual over time, the method should be modified (Bland & Altman, 2007). Only De Lombaert et al. (2023) did not use this method as the main aim of the study was not to evaluate accuracy.
Only four of the papers stated that the person collecting the measurements for the test method was blinded to the reference method (Cerejo et al. 2020; da Cunha et al., 2014; Binns et al., 1995; Caulkett et al., 1998), usually by having a separate person doing each. However, blinding is considered less important with an objective measurement such as blood pressure (BMJ Best Practice, 2023) than for studies requiring subjective interpretation, and it would not be possible to blind the interpretation of the data since it involves directly comparing the test and reference method. Only one paper using the Doppler method (Anjos et al., 2014) described that the person performing the measurements was experienced in the method and had received training. This could affect the results and interpretation since experience level has been shown to affect the coefficient of variability between measurements (Gouni et al., 2014), whereas oscillometry relies less on user skill or experience.
The ratio of limb/tail circumference to cuff size was shown to affect the accuracy of BP measurement (Cerejo et al., 2020) and the recommended ratio is 0.30–0.40. The range of ratios in the studies discussed was 0.30–0.60. Two studies (Martel et al., 2013; Zwijnenberg et al., 2011) did not calculate the ratio at all. Only six papers described specifically how they validated that the invasive method was accurate and reliable (Cerejo et al., 2017; Cerejo et al. 2020; da Cunha et al., 2014; Pedersen et al., 2002; Acierno et al., 2010; Cremer et al., 2020).
The sample size of five of these studies was smaller than the minimum recommended by ACVIM to validate a non-invasive device (Martel et al., 2013; De Lombaert et al., 2023; Branson et al., 1997; Pedersen et al., 2002; Zwijnenberg et al., 2011). da Cunha et al. (2014) excluded a patient from the analysis due to having outlier results, and it is not stated whether this removal affected the results, but this paper did have the largest sample size, so it is likely to be of less consequence than the other papers. Only one paper (Cremer et al., 2020) performed a power calculation, which showed the sample size was adequate to detect a significant difference between the methods.
Overall, although the study designs were individually sound and statistical analysis was suitable for the question asked, the results and conclusions from these papers are not in complete agreement. This may be due to the differences in methodology described above. However, there is more evidence for the accuracy of oscillometric measurements than for Doppler measurements. Recent work has shown that Doppler and oscillometry do not agree well with each other (Cerna et al., 2021). However, both are used widely by veterinary practices and veterinary staff may only have access to one method. In conclusion, this topic requires further research and the accuracy of non-invasive blood pressure remains a difficulty for the veterinary practitioner.
Methodology
Search Strategy
Databases searched and dates covered: |
CAB Abstracts at CABI Digital Library: 1977–2025 |
|---|---|
Search strategy: |
CAB Abstracts:
PubMed: (cat OR cats OR feline) AND (doppler or oscillometr*) AND (“blood pressure”) AND (invasive OR direct) |
Dates searches performed: |
21 May 2025 |
Exclusion / Inclusion Criteria
Exclusion: |
Species other than domestic cat, irrelevant to PICO question, not compared to invasive reference standard method. |
|---|---|
Inclusion: |
Compared to reference standard invasive telemetry, performed in domestic cats, oscillometric or Doppler measurement as test method. |
Search Outcome
Database |
Number of results |
Excluded – irrelevant to the PICO |
Excluded – not domestic cats |
Excluded – no comparison to reference standard invasive method |
Excluded – not available in English |
Excluded – full article not available |
Excluded – methodology and data not suitable for analysis |
Total relevant papers |
|---|---|---|---|---|---|---|---|---|
CAB Abstracts |
56 |
33 |
2 |
9 |
2 |
0 |
0 |
10 |
PubMed |
53 |
26 |
8 |
7 |
0 |
1 |
1 |
10 |
Citation tracking (cited by de Lombaert et al., 2023) |
2 |
0 |
0 |
0 |
0 |
0 |
0 |
2 |
Total relevant papers when duplicates removed |
14 |
|||||||
ORCiD
Fiona Douglas: https://orcid.org/0009-0002-0738-9285
Conflict of Interest
The authors declare no conflicts of interest.
References
- Acierno, M., Seaton, D., Mitchell, M. & da Cunha, A. (2010). Agreement between directly measured blood pressure and pressures obtained with three veterinary-specific oscillometric units in cats. Journal of the American Veterinary Medical Association. 237(4), 402–406. DOI: https://doi.org/10.2460/javma.237.4.402
- Anjos, T.M., Veado, J.C.C., Castro, M.C.N., De Araújo Diniz, S.A., Rocha, G.S.L., Da Silva, E.F., Araújo, C.A.V., Freitas, C.D., Maia, M.Q. & Tavares, C.A.P. (2014). Evaluation and comparison between methods of measurement of systolic blood pressure in healthy anesthetized cats. Brazilian Archive of Veterinary Medicine and Zootechnics. 66(4), 1051–1059. DOI: https://doi.org/10.1590/1678-6722 (Available translated at https://www.scielo.br/j/abmvz/a/SMCF8JJ4X8nCfncHm5sM4nj/?lang=pt (lAccessed: 20/03/2024).
- ANSI/AAMI SP10:2002. (2002). American Association of Medical Instrumentation, Standards for the performance of automated non-invasive blood pressure devices. Available at: https://mdcpp.com/doc/standard/ANSIAAMISP10-2002.pdf (Accessed 05 June 2026).
- Binns, S.H., Sisson, D.D., Buoscio, D.A. & Schaeffer, D.J. (1995). Doppler Ultrasonographic, Oscillometric Sphygmomanometric, and Photoplethysmographic Techniques for Noninvasive Blood Pressure Measurement in Anesthetized Cats. Journal of Veterinary Internal Medicine. 9(6), 405–414. DOI: https://doi.org/10.1111/j.1939-1676.1995.tb03301.x
- Bland, J.M. & Altman, D. (2007). Agreement Between Methods of Measurement with Multiple Observations Per Individual. Journal of Biopharmaceutical Statistics. 17(4), 571–582. DOI: https://doi.org/10.1080/10543400701329422
- BMJ Best Practice. (2024). Diagnostic test studies: assessment and critical appraisal. (Accessed: 20 March 2024).
- Branson, K. Wagner-Mann, C. & Mann, F. (1997). Evaluation of an oscillometric blood pressure monitor on anesthetized cats and the effect of cuff placement and fur on accuracy. Veterinary Surgery. 26(4), 347–353. DOI: https://doi.org/10.1111/j.1532-950x.1997.tb01510.x
- Brown, S., Atkins, C., Bagley, R., Carr, A., Cowgill, L., Davidson, M., Egner, B., Elliott, J., Henik, R., Labato, M., Littman, M., Polzin, D., Ross, L., Snyder, P. & Stepien, R. (2007). Guidelines for the Identification, Evaluation, and Management of Systemic Hypertension in Dogs and Cats. Journal of Veterinary Internal Medicine. 21(3), 542–558. DOI: https://doi.org/10.1111/j.1939-1676.2007.tb03005.x
- Caulkett, N.A., Cantwell, S.L. & Houston, D.M. (1998). A Comparison of Indirect Blood Pressure Monitoring Techniques in the Anesthetized Cat. Veterinary Surgery. 27(4), 370–377. DOI: https://doi.org/10.1111/j.1532-950x.1998.tb00143.x
- Cerejo, S.A., Teixeira-Neto, F.J., Garofalo, N.A., Rodrigues, J.C., Celeita-Rodríguez, N. & Lagos-Carvajal, A.P. (2017). Comparison of two species-specific oscillometric blood pressure monitors with direct blood pressure measurement in anesthetized cats. Journal of Veterinary Emergency and Critical Care. 27(4), 409–418. DOI: https://doi.org/10.1111/vec.12623
- Cerejo, S.A., Teixeira-Neto, F.J., Garofalo, N.A., Pimenta, E.L.M., Zanuzzo, F.S. & Klein, A.V. (2020). Effects of cuff size and position on the agreement between arterial blood pressure measured by Doppler ultrasound and through a dorsal pedal artery catheter in anesthetized cats. Veterinary Anaesthesia and Analgesia. 47(2), 191–199. DOI: https://doi.org/10.1016/j.vaa.2019.11.001
- Cerna, P. Archontakis, P.E., Cheuk, H.O. & Gunn-Moore, D.A. (2021). Comparison of Doppler ultrasonic and oscillometric devices (with or without proprietary optimisations) for non-invasive blood pressure measurement in conscious cats. Journal of Feline Medicine and Surgery. 23(2), 121–130. DOI: https://doi.org/10.1177/1098612X20932407
- Cremer, J., da Cunha, A., Aulakh, K., Liu, C-C. & Acierno, M.J. (2020). Validation of the oscillometric blood pressure monitor Vet20 Suntech in anesthetized healthy cats. Veterinary Anaesthesia and Analgesia. 47(3), 309–314. DOI: https://doi.org/10.1016/j.vaa.2019.12.007
- da Cunha, A.F., Saile, K., Beaufrère, H., Wolfson, W., Seaton, D. & Acierno, M.J. (2014). Measuring level of agreement between values obtained by directly measured blood pressure and ultrasonic Doppler flow detector in cats. Journal of Veterinary Emergency and Critical Care. 24(3), 272–278. DOI: https://doi.org/10.1111/vec.12161
- De Lombaert, M.C.M., Lourenço, B.N., Coleman, A.E., Arne, A.M., Berghaus, R.D. & Schmiedt, C.W. (2023). Effect of gabapentin on ambulatory, direct, systemic arterial blood pressure in apparently healthy cats in the at-home and in-clinic environments. Journal of Feline Medicine and Surgery. 25(8). DOI: https://doi.org/10.1177/1098612X231188770
- Garofalo, N.A., Teixeira, N., Alvaides, R.K., de Oliveira, F.A., Pignaton, W. & Pinheiro, R.T. (2012) Agreement between direct, oscillometric and Doppler ultrasound blood pressures using three different cuff positions in anesthetized dogs. Veterinary Anaesthesia and Analgesia. 39(4), 324–334. DOI: https://doi.org/10.1111/j.1467-2995.2012.00711.x
- Giavarina, D. (2015). Understanding Bland Altman analysis. Biochemia Medica (Zagreb). 25(2), 141– DOI: https://doi.org/10.11613/BM.2015.015
- Gouni, V., Tissier, R., Misbach, C., Balouka, D., Bueno, H., Pouchelon, J-L., Lefebvre, H.P. & Chetboul, V. (2014). Influence of the observer’s level of experience on systolic and diastolic arterial blood pressure measurements using Doppler ultrasonography in healthy conscious cats. Journal of Feline Medicine and Surgery. 17(2), 94–100. DOI: https://doi.org/10.1177/1098612x14532087
- Grubb, T., Sager, J., Gaynor, J.S., Montgomery, E., Parker, J.A., Shafford, H. & Tearney, C. (2020). 2020 AAHA Anesthesia and Monitoring Guidelines for Dogs and Cats. Journal of the American Animal Hospital Association. 56(2), 59–82. DOI: https://doi.org/10.5326/JAAHA-MS-7055
- Haberman, C.E., Morgan, J.D., Kang, C.W. & Brown, S.A. (2006). Evaluation of Doppler ultrasonic and oscillometric methods of indirect blood pressure measurement in cats. International Journal of Applied Research in Veterinary Medicine. 2(4), 279–289.
- Howick, J., Phillips, B., Ball, C., Badenoch., D., Straus, S., Haynes, B. & Dawes, M. (2009). Oxford Centre for Evidence-Based Medicine: Levels of Evidence (March 2009). Centre for Evidence-Based Medicine. Available at: https://www.cebm.ox.ac.uk/resources/levels-of-evidence/oxford-centre-for-evidence-based-medicine-levels-of-evidence-march-2009 (Accessed: 5 July 2023)
- Jepson, E.R. (2011). Feline Systemic Hypertension: Classification and pathogenesis. Journal of Feline Medicine and Surgery. 13(1), 25–34. DOI: https://doi.org/10.1016/j.jfms.2010.11.007
- Martel, E., Egner, B., Brown, S.A., King, J.N., Laveissiere, A., Champeroux, P. & Richard, S. (2013). Comparison of high-definition oscillometry — a non-invasive technology for arterial blood pressure measurement — with a direct invasive method using radio-telemetry in awake healthy cats. Journal of Feline Medicine and Surgery. 15(12), 1104–1113. DOI: https://doi.org/10.1177/1098612X13495025
- Midon, M., Yamada, D.I., Zangirolami Filho, D., Simionato, G.C., Feringer Junior, W.H., Conde, G. & Hofmeister, E.H. (2023). Comparison of Four Arteries for Invasive Blood Pressure Measurements and Dixtal 2010 Oscillometric Values in Horses Anesthetized With Isoflurane: Does the Artery Matter? Journal of Equine Veterinary Science. 121, 104219. DOI: https://doi.org/10.1016/j.jevs.2023.104219
- Monteiro, E.R., Campagnol, D., Bajotto, G.C., Simões, C.R. & Rassele, A.C. (2013). Effects of 8 hemodynamic conditions on direct blood pressure values obtained simultaneously from the carotid, femoral and dorsal pedal arteries in dogs. Journal of Veterinary Cardiology. 15(4), 263–270. DOI: https://doi.org/10.1016/j.jvc.2013.07.002
- Navarro, I., Summers, S., Rishniw, M. & Quimby, J. (2022). Cross-sectional survey of non-invasive indirect blood pressure measurement practices in cats by veterinarians. Journal of Feline Medicine and Surgery. 24(12), 1195-1202.DOI: https://doi.org/10.1177/1098612X211067015
- Pedersen, K.M., Butler, M.A., Ersbøll, A.K. & Pedersen, H.D. (2002). Evaluation of an oscillometric blood pressure monitor for use in anesthetized cats. Journal of the American Veterinary Medical Association. 221(5), 646–650. DOI: https://doi.org/10.2460/javma.2002.221.646
- Skelding, A. & Valverde, A. (2020). Non-invasive blood pressure measurement in animals: Part 1 - Techniques for measurement and validation of non-invasive devices. The Canadian Veterinary Journal. 61(4), 368–374.
- Stergiou, G., Alpert, B., Mieke, S., Asmar, R., Atkins, N., Eckert, S., Frick, G., Friedman, B., Graßl, T., Ichikawa, T., Ioannidis, J., Lacy, P., McManus, R., Murray, A., Myers, M., Palatini, P., Parati, G., Quinn, D., Sarkis, J., Shennan, A., Usuda, T., Wang, J., Wu, C. & O’Brien, E. (2018). A universal standard for the validation of blood pressure measuring devices. Journal of Hypertension. 36(3), 472–478. DOI: https://doi.org/10.1097/hjh.0000000000001634
- Taylor, S.S., Sparkes, A.H., Briscoe, K., Carter, J., Sala, S.C., Jepson, R.E., Reynolds, B. & Scansen, B.A. (2017). ISFM Consensus Guidelines on the Diagnosis and Management of Hypertension in Cats. Journal of Feline Medicine and Surgery. 19(3), 288–303. DOI: https://doi.org/10.1177/1098612x17693500
- Troìa, R., Mascalzoni, G., Calipa, S., Magagnoli, I., Dondi, F. & Giunti, M.. (2019). Multiorgan dysfunction syndrome in feline sepsis: prevalence and prognostic implication. Journal of Feline Medicine and Surgery, 21(6), 559–565. DOI: https://doi.org/10.1177/1098612X18792106
- Zwijnenberg, R.J., del Rio, C.L., Cobb, R.M., Ueyama, Y. & Muir, W.W. (2011). Evaluation of oscillometric and vascular access port arterial blood pressure measurement techniques versus implanted telemetry in anesthetized cats. American Journal of Veterinary Research. 72(8), 1015–1021. DOI: https://doi.org/10.2460/ajvr.72.8.1015
Contribute to the evidence
There are two main ways you can contribute to the evidence base while you are enhancing your CPD:
- Tell us your information need
- Write a Knowledge Summary
Either way, you will be helping to add to the evidence base, and strengthen the decisions that veterinary professionals around the world make to give animals the best possible care. Learn more here: https://veterinaryevidence.org/index.php/ve/author-hub
Licence
Copyright (c) 2026 Fiona Douglas
This work is licensed under a Creative Commons Attribution 4.0 International License.
Veterinary Evidence uses the Creative Commons copyright Creative Commons Attribution 4.0 International License. That means users are free to copy and redistribute the material in any medium or format. Remix, transform, and build upon the material for any purpose, even commercially – with the appropriate citation.
Intellectual Property Rights
Authors of Knowledge Summaries submitted to RCVS Knowledge for publication will retain copyright in their work, and will be required to grant to RCVS Knowledge a non-exclusive licence to publish including but not limited to the right to publish, re-publish, transmit, sell, distribute and otherwise use the materials in all languages and all media throughout the world, and to licence or permit others to do so.
Disclaimer
Knowledge Summaries are a peer-reviewed article type which aims to answer a clinical question based on the best available current evidence. It does not override the responsibility of the practitioner. Informed decisions should be made by considering such factors as individual clinical expertise and judgement along with patient’s circumstances and owners’ values. Knowledge Summaries are a resource to help inform and any opinions expressed within the Knowledge Summaries are the author's own and do not necessarily reflect the view of the RCVS Knowledge. Authors are responsible for the accuracy of the content. While the Editor and Publisher believe that all content herein are in accord with current recommendations and practice at the time of publication, they accept no legal responsibility for any errors or omissions, and make no warranty, express or implied, with respect to material contained within. For further information please refer to our Terms of Use.